700 EX Nano Gas Chromatograph

700 EX Nano Gas Chromatograph

Introduction

The 700 EX Nano GC gas chromatograph is specifically designed for industrial analysis, making it suitable for online or portable applications in explosion-proof industrial settings. The instrument
adopts a compact explosion-proof design, is compact and lightweight, and can continuously and accurately monitor component concentrations in sample fluids in applications such as natural gas, refining, chemicals, power, steel, and environmental applications.

The 700 EX Nano GC gas chromatograph can be installed directly on pipelines or walls, eliminating the need for expensive analysis booths; simple pretreatment of sample gases enables online analysis of various gas components.

Thanks to extremely low gas consumption, a 40L carrier cylinder can provide continuous use for the instrument for up to 7 years; With a minimum operating power of up to 6 W, it can be equipped with a solar power system, suitable for outdoor or grid-free scenarios.

Features

> Compact size, lightweight

> Short analysis time, improves production efficiency
> Can be built into multiple pressure regulators and flow selection valves
> Pipeline or wall mounting, can be installed outdoors
> Local interface sensing touch or gesture operation
> Very low carrier and sample gas consumption

> Extremely low power consumption

> Low procurement, installation, and maintenance costs

Specifications

Injection typeMiniature injectors (fixed FX, straight-through VV, reverse blow BF), sample valves
Chromatographic columnCapillary columns, micro-filled columns
Temperature controlMaximum temperature is 250°C, with a control accuracy of 0.01°C
Pressure controlElectronic pressure control, control accuracy of 0.01 psi
Detection channelsUp to 2 (depending on application)
DetectorThermal conductivity detector (TCD), minimum detection limit 0.5 ppm
Flame ionization detector (FID), minimum detection limit 0.05 ppm
Concentration range0 – 100%
Analyze the time≤ 2 min (Typical)
Repeatability≤ 0.1% RSD (Controlled Environment)
Carrier gas typeNitrogen N2, argon He, helium He, hydrogen H2, air
Carrier gas consumption2 mL/min (typical channel value)
Air Connection Interface1/8″ Swagelock
Data storageSQL databases and file servers
WorkstationNextWL, EliteChrom
Operation terminal4.3″ LCD color screen, touch sensing keys, gesture control (optional), external computer
Sample injection channelInternal or external sample flow path switching (optional)
Communication interfaceNetwork port RJ45, serial port RS232/RS485, wireless (optional)
Communication protocolsModbusRTU、ModbusTCP、GRPC
Protection ratingIP66
Explosion-proof levelEx db IIC T6 Gb
Instrument dimensionsΦ170×430 mm
Φ170×380 mm (no local interface configuration)
Installation methodWalls or pipes
Air sourceGas cylinders or gas generators
Weight9 kg
Power supply12 VDC, minimum 6 W (depending on application)
Work environmentTemperature: -20 – 55 °C, Humidity: 0 – 95%

Applications

Petroleum and natural gas exploration
Mud logging
Permanent gas,
Refinery gas,
Liquefied petroleum gas
Sspecialty gas quality control
Syngas and catalytic reactions
Biogas
Cracking, ethylene
Power plant exhaust gas
Reforming and catalytic processes
Isomerization devices
Coal-to-olefin
Flares, exhaust gas
Fermentation process
Gas desulfurization
Natural gas pipeline monitoring
Natural gas calorific value analysis
Natural gas odorizers
Coke oven gas,
Blast furnace gas
Sulfide gas
Ambient gases